In the Mira hormone monitor device firmware v1.7.1.47 build 01070147, a remote unauthenticated attacker within BLE range (approximately 10–30 meters) can silently rebind the device to an attacker-con…
High CVSS 8.8
Summary
In the Mira hormone monitor device firmware v1.7.1.47 build 01070147, a remote unauthenticated attacker within BLE range (approximately 10–30 meters) can silently rebind the device to an attacker-controlled account, extract stored hormone measurements in cleartext, cause a denial-of-service via malformed or undocumented command opcodes, and passively track the user via a static random BLE address that never rotates.
In-depth triage
No in-depth report has been generated yet (DR-003 v2 AI pipeline is under construction).
Sources
- NVD DATABASE
Original Links
- https://github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2026/icsma-26-223-01.json
- https://www.cisa.gov/news-events/ics-medical-advisories/icsma-26-223-01
Timeline
- nvd_ingest NVD